START_INST_TAB  = P_FC1_C2
VERSION         = V00.02
DATE            = 92/09/04
AUTHOR          = I. Heinrichsen
COMMENT
COMMENT ***************************************************************
COMMENT
COMMENT                        P_FC1_C2 
COMMENT
COMMENT  Provides the transformation factors between power in W/pixel on 
COMMENT  the C200 detector and the electrical power to be applied to  
COMMENT  the fine calibration source FCS1 for this flux. 
COMMENT  This relation is given for each C200 filter independently.
COMMENT  In addition the command mnemonic for the correct selection of the 
COMMENT  thermal radiation source (trs) mode is given.
COMMENT
COMMENT ***************************************************************
COMMENT
NUM_OF_KEYS     = 11
NUM_OF_ROWS     = 33
COMMENT --------------------------------------------------------
KEY_NAME01      = FLUX
COMMENT         Flux in W/pixel on the C200 detector
KEY_FORM01      = F
COMMENT --------------------------------------------------------
KEY_NAME02      = ELECT_POWER_9
COMMENT         Electrical power in [mW] to be applied to FCS1 for filter
COMMENT         C200_160_UM 
KEY_FORM02      = F
COMMENT --------------------------------------------------------
KEY_NAME03      = TRS_MODE_9
COMMENT         trs mode mnemonic to be used for the given flux for filter
COMMENT         C200_160_UM
KEY_FORM03      = C
COMMENT --------------------------------------------------------
KEY_NAME04      = ELECT_POWER_10
COMMENT         Electrical power in [mW] to be applied to FCS1 for filter
COMMENT         C200_200_UM 
KEY_FORM04      = F
COMMENT --------------------------------------------------------
KEY_NAME05      = TRS_MODE_10
COMMENT         trs mode mnemonic to be used for the given flux for filter
COMMENT         C200_200_UM
KEY_FORM05      = C
COMMENT --------------------------------------------------------
KEY_NAME06      = ELECT_POWER_11
COMMENT         Electrical power in [mW] to be applied to FCS1 for filter
COMMENT         C200_180_UM
KEY_FORM06      = F
COMMENT --------------------------------------------------------
KEY_NAME07      = TRS_MODE_11
COMMENT         trs mode mnemonic to be used for the given flux for filter
COMMENT         C200_180_UM
KEY_FORM07      = C
COMMENT --------------------------------------------------------
KEY_NAME08      = ELECT_POWER_12
COMMENT         Electrical power in [mW] to be applied to FCS1 for filter
COMMENT         C200_135_UM
KEY_FORM08      = F
COMMENT --------------------------------------------------------
KEY_NAME09      = TRS_MODE_12
COMMENT         trs mode mnemonic to be used for the given flux for filter
COMMENT         C200_135_UM
KEY_FORM09      = C
COMMENT --------------------------------------------------------
KEY_NAME10      = ELECT_POWER_13
COMMENT         Electrical power in [mW] to be applied to FCS1 for filter
COMMENT         C200_120_UM
KEY_FORM10      = F
COMMENT --------------------------------------------------------
KEY_NAME11      = TRS_MODE_13
COMMENT         trs mode mnemonic to be used for the given flux for filter
COMMENT         C200_120_UM
KEY_FORM11      = C
COMMENT 
COMMENT -------------------------------------------------------------------
COMMENT               160         200        180        135        120   
COMMENT
COMMENT FLUX[W/pix]  9    M9    10   M10   11   M11   12   M12   13   M13
COMMENT -------------------------------------------------------------------
1.00E-18  0.03 TRS2_ON  0.03 TRS2_ON  0.03 TRS2_ON  0.04 TRS2_ON  0.09 TRS2_ON
1.78E-18  0.04 TRS2_ON  0.04 TRS2_ON  0.04 TRS2_ON  0.05 TRS2_ON  0.10 TRS2_ON
3.16E-18  0.04 TRS2_ON  0.05 TRS2_ON  0.05 TRS2_ON  0.06 TRS2_ON  0.11 TRS2_ON
5.62E-18  0.05 TRS2_ON  0.06 TRS2_ON  0.06 TRS2_ON  0.06 TRS2_ON  0.13 TRS2_ON
1.00E-17  0.06 TRS2_ON  0.07 TRS2_ON  0.07 TRS2_ON  0.08 TRS2_ON  0.14 TRS2_ON
1.78E-17  0.07 TRS2_ON  0.08 TRS2_ON  0.08 TRS2_ON  0.09 TRS2_ON  0.16 TRS2_ON
3.16E-17  0.08 TRS2_ON  0.10 TRS2_ON  0.09 TRS2_ON  0.10 TRS2_ON  0.18 TRS2_ON
5.62E-17  0.09 TRS2_ON  0.12 TRS2_ON  0.11 TRS2_ON  0.12 TRS2_ON  0.21 TRS2_ON
1.00E-16  0.11 TRS2_ON  0.14 TRS2_ON  0.13 TRS2_ON  0.14 TRS2_ON  0.24 TRS2_ON
1.78E-16  0.13 TRS2_ON  0.18 TRS2_ON  0.15 TRS2_ON  0.16 TRS2_ON  0.28 TRS2_ON
3.16E-16  0.15 TRS2_ON  0.22 TRS2_ON  0.18 TRS2_ON  0.19 TRS2_ON  0.32 TRS2_ON
5.62E-16  0.18 TRS2_ON  0.27 TRS2_ON  0.22 TRS2_ON  0.23 TRS2_ON  0.37 TRS2_ON
1.00E-15  0.22 TRS2_ON  0.34 TRS2_ON  0.27 TRS2_ON  0.27 TRS2_ON  0.43 TRS2_ON
1.78E-15  0.27 TRS2_ON  0.45 TRS2_ON  0.34 TRS2_ON  0.33 TRS2_ON  0.51 TRS2_ON
3.16E-15  0.33 TRS2_ON  0.59 TRS2_ON  0.42 TRS2_ON  0.40 TRS2_ON  0.61 TRS2_ON
5.62E-15  0.41 TRS2_ON  0.80 TRS2_ON  0.54 TRS2_ON  0.49 TRS2_ON  0.74 TRS2_ON
1.00E-14  0.52 TRS2_ON  1.13 TRS2_ON  0.70 TRS2_ON  0.61 TRS2_ON  0.91 TRS2_ON
1.78E-14  0.66 TRS2_ON  1.68 TRS2_ON  0.95 TRS2_ON  0.78 TRS2_ON  1.13 TRS2_ON
3.16E-14  0.86 TRS2_ON  2.63 TRS2_ON  1.31 TRS2_ON  1.01 TRS2_ON  1.43 TRS2_ON
5.62E-14  1.16 TRS2_ON  4.44 TRS2_ON  1.90 TRS2_ON  1.34 TRS2_ON  1.86 TRS2_ON
1.00E-13  1.62 TRS2_ON  8.15 TRS2_ON  2.91 TRS2_ON  1.85 TRS2_ON  2.50 TRS2_ON
1.78E-13  2.35 TRS2_ON 16.38 TRS2_ON  4.78 TRS2_ON  2.66 TRS2_ON  3.47 TRS2_ON
3.16E-13  3.59 TRS2_ON 35.82 TRS2_ON  8.47 TRS2_ON  4.02 TRS2_ON  5.06 TRS2_ON
5.62E-13  5.88 TRS2_ON 49.99 TRS2_ON 16.47 TRS2_ON  6.51 TRS2_ON  7.82 TRS2_ON
1.00E-12 10.42 TRS2_ON 49.99 TRS2_ON 35.10 TRS2_ON 11.42 TRS2_ON 12.99 TRS2_ON
1.78E-12 20.22 TRS2_ON 49.99 TRS2_ON 49.99 TRS2_ON 21.92 TRS2_ON 23.53 TRS2_ON
3.16E-12 42.85 TRS2_ON 49.99 TRS2_ON 49.99 TRS2_ON 46.01 TRS2_ON 46.61 TRS2_ON
5.62E-12 49.99 TRS2_ON 49.99 TRS2_ON 49.99 TRS2_ON 49.99 TRS2_ON 49.99 TRS2_ON
1.00E-11 49.99 TRS2_ON 49.99 TRS2_ON 49.99 TRS2_ON 49.99 TRS2_ON 49.99 TRS2_ON
1.78E-11 49.99 TRS2_ON 49.99 TRS2_ON 49.99 TRS2_ON 49.99 TRS2_ON 49.99 TRS2_ON
3.16E-11 49.99 TRS2_ON 49.99 TRS2_ON 49.99 TRS2_ON 49.99 TRS2_ON 49.99 TRS2_ON
5.62E-11 49.99 TRS2_ON 49.99 TRS2_ON 49.99 TRS2_ON 49.99 TRS2_ON 49.99 TRS2_ON
1.00E-10 49.99 TRS2_ON 49.99 TRS2_ON 49.99 TRS2_ON 49.99 TRS2_ON 49.99 TRS2_ON
END_INST_TAB  


############################################################################
#
# Effective Revision History for  P_FC1_C2
# Author: Luciano Orsini
# Creation Date: 09/03/94
#
# Version: $Id: p_fc1_c2.txt,v 1.3 1994/11/15 08:35:19 ittlib Exp $
#

#
# Revision History: $Log: p_fc1_c2.txt,v $
# Revision 1.3  1994/11/15  08:35:19  ittlib
# Update to logic to version 2.3
#
# Revision 3.3  1994/09/28  13:51:03  lorsini
# 25/07/1994: revised FM data H. Krueger
# A set of FCS tables was generated to support the EE1 test. The new
# tables contain considerably improved data for the behaviour of the FCS
# of PHT. An individual fit for all 4 TRSs were performed and the data
# were fitted for every filter independently.
#
# Revision 3.2  1994/07/19  13:07:17  lorsini
# 18/07/1994: version for EE1 test, FM data. I. Heinrichsen
#
# Revision 3.1  1994/03/15  09:35:36  lorsini
# Development for EE1 test ( Initial Version ).
#
#