Table B.1:
Description of 16 bits ERD bit flag variable PPERPIXF. LSB = 0, MSB = 15.
| bit range |
item |
decimal |
meaning |
| |
|
value |
|
| 0 |
auto data reduction |
0 |
not active; all telemetry (TM) blocks can be |
| |
|
|
transferred |
| |
|
1 |
active; one or more TM blocks of the |
| |
|
|
measurement could not be transmitted. |
| |
|
|
science data shall not be used, |
| |
|
|
they may be corrupted. |
| 1-2 |
not used |
- |
none |
| 3-4 |
selected TRS of FCS1 |
0 |
both TRSs not connected |
| |
|
1 |
only TRS1 connected |
| |
|
2 |
only TRS2 connected |
| 5-6 |
selected TRS of FCS2 |
0 |
both TRSs not connected |
| |
|
1 |
only TRS1 connected |
| |
|
2 |
only TRS2 connected |
| 7 |
SW override |
0 |
not active |
| |
|
1 |
active; no operational but purely diagnostic |
| |
|
|
set-up |
| 8-11 |
TM data structure |
0 |
P1 data structure |
| |
|
1 |
P2 data structure |
| |
|
2 |
P3 data structure |
| |
|
3 |
PHT-S data structure |
| |
|
4,5 |
not used |
| |
|
6 |
C100 data structure |
| |
|
7 |
C200 data structure |
| |
|
8..12 |
not used |
| |
|
13 |
diagnostic data structure |
| |
|
14 |
extended house-keeping (HK) data structure |
| |
|
15 |
not used |
| 12-15 |
integration time |
0..15 |
selected fundamental integration time |
| |
|
|
(reset interval) |
| |
|
|
exponent n in 2(7-n) [s] |
| |
|
|
0 n 15 |