next up previous contents index
Next: 5.6 Photometric Accuracy Up: 5. Calibration and Performance Previous: 5.4 Dark Current Determination


5.5 In-orbit Sensitivity of the LWS - Detector Performance

The sensitivity of the LWS is based on the knowledge of the noise equivalent power (NEP), which is defined for each detector as the noise obtained in dark current measurements with 0.5 s integration time divided by the detector responsivity $S$ in A/W.


\begin{displaymath}
NEP\,=\,noise\,/S \,~~~~~~~[{\rm W/Hz^{-1/2}}]
\end{displaymath} (5.1)

The noise figures are based on a measurement of the noise under dark conditions taken in revolution 650 (see Section 5.4). To obtain a noise figure 50 minutes of data were taken with the satellite pointing at a dark region of the sky and with a Fabry-Pérot in the beam with its etalons set non-parallel. The noise was estimated from the standard deviation of a Gaussian curve fitted to a histogram of the photocurrent values after deglitching. As most of the LWS observations, they have been performed using 0.5s reset times, which is equivalent to a 1Hz bandwidth.

To avoid making asumptions about the transmission of the telescope and instrument the LWS detector responsivity is not measured directly. Instead the instrument response is derived from the calibration source (Uranus) and its associated model spectrum; it is given as $S_i(\nu)$ [Acm$^2\,\mu$mW$^{-1}$]. We can calculate the Noise Equivalent Spectral Density (NESD) directly from this value:


\begin{displaymath}
NESD(\nu)\,=\,noise\,/S_i(\nu) ~~~~~~~~ [{\rm W\,cm^{-2}\,
\mu m^{-1}\, Hz^{-1/2}}]
\end{displaymath} (5.2)

To convert this to NEP of the detectors requires knowledge of the telescope effective area and the instrument throughput. The instrument throughput was not directly measured on the ground and is complex to calculate. We therefore refer everything to the entrance aperture of the instrument by assuming knowledge of the instrument spectral resolution in grating mode and the effective area of the ISO telescope as a function of wavelength. This effective area has been calculated from an optical model of the ISO telescope and is given for the central wavelength of each detector in Table 5.5. The spectral resolution of the grating is measured from narrow emission line spectra. The calculated instrument NESD and NEFD (noise equivalent flux density) figures and the estimated instrument NEP are given in Table 5.5.


Table 5.5: In-orbit sensitivity of the LWS instrument.
Detector Centre Noise for NESD Telescope NEP $\eta\tau$ cross over
  $\lambda$ 0.5 s resets [10$^{-19}$Wcm$^{-2}$ eff. area [10$^{-16}$   flux
  [$\mu $m] [10$^{-17}$A] $\mu $m$^{-1}$Hz$^{-1/2}$] [cm$^2$] WHz$^{-1/2}$]   [Jy]
SW1 46.13 4.17 4.43 2460 3.16 0.0046 260
SW2 56.11 3.83 7.75 2431 5.47 0.0046 1460
SW3 66.03 1.91 3.00 2346 2.04 0.0095 710
SW4 75.61 2.98 5.40 2262 3.55 0.0044 1530
SW5 84.68 2.13 4.28 2205 2.74 0.0044 1330
LW1 102.25 2.68 0.849 1922 0.980 0.0095 360
LW2 122.04 2.26 0.231 1838 0.254 0.0095 42
LW3 141.63 3.19 0.122 1838 0.135 0.0142 28
LW4 160.38 3.83 0.202 1753 0.212 0.0108 80
LW5 177.74 2.21 0.691 1583 0.656 0.0033 360

The instrument NEPs increased by an average of a factor of four compared to pre-launch values. A factor of two increase was expected as 0.5 s amplifier resets were used in-orbit, whereas in the ground testing the NEP was measured using 2 s resets and the noise is proportional to the square root of the reset time for resets up to about 4 s (Shaver et al. 1983, [38]). The increased NEP was also due to the decreased responsivity which many of the detectors exhibited in-orbit, some of which was due to the decrease in detector bias voltage (to reduce spontaneous spiking) and the effects of ionising radiation on the detectors. For further analysis of the detector performance in-orbit compared to that on the ground see Leeks et al. 2001, [25].

If the power $P$ falling on the detectors is low enough so that the noise is dominated by the detector read noise, then the signal-to-noise ratio $\sigma$ expected in an integration time $T$ is given by:


\begin{displaymath}
\sigma = \frac{P}{NEP} \sqrt{2T}
\end{displaymath} (5.3)

In teh pther extreme, when the power falling on the detectors is high enough such that the noise is dominated by the shot noise in the photon stream, the signal-to-noise $\sigma$ is given by:


\begin{displaymath}
\sigma = (\frac{P}{4h\nu/\eta\tau})^{0.5} \sqrt{2T}
\end{displaymath} (5.4)

The power at which the cross over between read noise and photon noise occurs has been derived in a study of the signal-to-noise ratio, performed using internal illuminators measurements (see Swinyard et al. 2000, [41] for more details). They are listed in Table 5.5.


next up previous contents index
Next: 5.6 Photometric Accuracy Up: 5. Calibration and Performance Previous: 5.4 Dark Current Determination
ISO Handbook Volume III (LWS), Version 2.1, SAI/1999-057/Dc