Next: 5.3 Astronomical Calibration procedure
Up: 5 Calibration and Accuracy
Previous: 5.1 Calibration philosophy
- Debiasing parameters
- These are measured with the
internal illuminators, using different compensation voltages at the
start of the ramps. The compensation levels induce different
effective bias levels for the ramps. From the second order polynomial
fits to the data the debiasing parameter a (see section 4.2)
can be determined.
- Detector Dark Current
- The dark currents can only be
measured together with any current induced by background or stray
radiation. To minimize this background radiation
the measurement is done by placing a Fabry-Perot - deliberately made
non-parallel - in the beam. Tests have shown that
this gives the lowest background on the detectors, the FP acting as a
highly efficient blocking filter. ISO is then pointed to a
relatively clear part of the sky to get minimal straylight from
off-axis sources. The observation
is done with several bias levels. The grating is scanned to look for
spectral signatures in the dark current, and the illuminators are used
to check for responsivity changes during the dark current measurement.
No need for in-orbit characterisation could be identified. All characteristics
are determined in the laboratory and while building the instrument. The
important characteristics are:
- The time constants of the high pass electronic filters
- The gains of the amplifiers
During the Spacecraft Commissioning Phase (SCP) several tests were
done to detect possible large changes in these values that would
point to problems with the instrument.
The optimum offsets for the
Fabry-Perot etalons for FPS and FPL are determined by observing an unresolved
line at several offset values around the
expected optimum values (offset mapping). This gives a set of line
spectra. For each of these spectra the FWHM of the line and the peak heights
are determined. From the contour maps of the FWHM of the line versus the
two FP offset values (for the coils of one FP) the position of the minimum
FWHM is determined. The offset values corresponding to this minimum
are optimum for parallelism.
The beam pattern of the instrument is determined by taking full grating spectra
over a raster map of a strong source. The result is a map of the beam-pattern
of the instrument as a function of wavelength.
Next: 5.3 Astronomical Calibration procedure
Up: 5 Calibration and Accuracy
Previous: 5.1 Calibration philosophy
N.Trams, ISO Science Operations Team
Using inputs from:
C.Gry, T. Lim, LWS Instrument Dedicated Team
A.Harwood, P.E.Clegg, B.Swinyard, K.King, LWS Instrument Team
S.Lord, S.Unger, IPAC.