For version 6 of the pipeline, a new wavelength calibration was derived consisting of
The size of the grating spectral element was determined and found to be equal to the size that was quoted in the LWS Observers Manual, i.e. 0.29 m for the SW detectors and 0.60 m for the LW detectors.
The sources and lines that were used are given in Table 5.2.
Source | Type | Lines |
Grating wavelength calibration | ||
NGC6543 | PN | 52, 57, 63, 88 m |
NGC6826 | PN | 52, 57, 63, 88 m |
G298.228 | HII region | 52, 57, 63, 88, 145, 158 m |
IRAS15408-5356 | HII region | 52, 57, 63, 88, 145, 158 m |
FP wavelength calibration | ||
NGC7027 | PN | 63.184 [OI], 157.741 [CII] |
NGC3603 P1 | HII region | 51.814 [OIII], 57.317 [NIII], 157.741 [CII] |
NGC6826 | PN | 51.814 [OIII], 88.356 [OIII], 157.741 [CII] |
G0.6-0.6 | HII region | 51.814 [OIII], 57.317 [NIII], 63.184 [OI], 121.898 [NII] |
145.525 [OI], 157.741 [CII] |
The accuracy for the grating wavelength calibration quoted here is based on several observations of NGC6543, and two targets that were not used for the derivation of the coefficients: NGC7027 and W Hya. The measured lines had intensities ranging between and W/cm . 85% of the measured lines show a deviation from their expected wavelengths of less than 25% of a grating resolution element (0.07 m for the SW detectors and 0.15 m for the LW detectors). The maximum deviation that was measured was 0.27 m on LW4 or 45% of a resolution element. See also section 7.5 for some caveats on the grating wavelength calibration. The accuracy that is achieved for the wavelength with the grating is given in Table 5.3.
mode | Accuracy |
grating calibration | 25% of a resol. element |
0.07 m for SW detectors | |
0.15 m for LW detectors | |
FPS calibration | m RMS |
scatter in the fitted calibration | |
FPL calibration | m RMS |
scatter in the fitted calibration | |
FP absolute accuracy | 40 km/s due to errors in FPS |
and FPL calibration |
The RMS scatter in the fitted calibration of the Fabry-Perots gives a measure of the internal accuracy of the calibration. This was determined to be m for FPS and m for FPL. However, tests using several strong lines on a single source have shown a significant velocity difference between lines observed with FPS and FPL. The differences can be as large as 80 km/sec with FPL giving systematically higher velocities. This difference is due to an error in both the FPS and the FPL calibrations (see section 7.7).
The sources and lines that were used are given in Table 5.2. The accuracy that is achieved for the wavelength with the FPs is given in Table 5.3.
The accuracy of the photometric calibration is determined by a number of factors:
All these factors together lead to a photometric accuracy for LWS grating mode spectra of 10% between repeated scans on the same detector (this is mainly due to the effect of responsivity changes), and 30% between adjacent detectors (mainly due to responsivity changes and dark backgroound removal problems).
For the Fabry-Perot mode, the photometric accuracy was determined by comparing the integrated line fluxes observed with the FP with the fluxes observed with the grating or line fluxes published in the literature. The sources and lines are given in Table 5.4. It was found that for strong lines accuarcy is typically better than 30%. For faint lines however, the FP fluxes can be off by almost a factor 2. This is mainly due to the removal of the darkcurrent and the responsivity drift correction which are known to be less accurate for low signal levels (see also section 7.13).
Table 5.4: The sources
and lines in this table were used for the determination of the photometris
accuarcy of the Fabry-Perot data.