7.3 Last ramp in format

 

Due to a timing problem, the last ramp of every format (2 second period, i.e. in most cases every fourth ramp), will have a larger value for the photocurrent than expected. This is most clearly seen for strong sources, but could cause an increase in noise in all circumstances. The solution that will be implemented in version 7.0 of the OLP software will be to remove the last point of every ramp.



N.Trams, ISO Science Operations Team
Using inputs from:
C.Gry, T. Lim, LWS Instrument Dedicated Team
A.Harwood, P.E.Clegg, B.Swinyard, K.King, LWS Instrument Team
S.Lord, S.Unger, IPAC.