 
 
 
 
 
 
 
 
 
 
 Next: 4.2 AOT 1
Up: 4. AOT overview
 Previous: 4. AOT overview
 
Subsections
4.1 AOT Introduction
Astronomical observations with ISO are carried out via observing modes known as 
AOTs, Astronomical Observation Templates. These are modes in which 
the basic operating parameters for the instruments are predefined (e.g. which 
sub-instrument to use at what wavelength resolution), leaving the user to 
select wavelengths to scan between, enter flux levels etc.
There are two main observing modes with the SWS.
- 1.
- Grating-Only Observations (Medium spectral resolution). The spectral region of interest is directed to an array of 12 
detectors. The spectral range instantaneously covered by the array is about 8 
grating resolution elements, with gaps between the individual detectors
 (except for Band 3). For an observation, the 
grating is scanned in small steps to fully sample
the desired wavelength range and provide a certain redundancy in case of
detector failures or detector memory effects. Different
scan schemes are used to obtain high fidelity line profiles, scans of
wavelength ranges, and quick low-resolution
full scans of the SWS wavelength range.
- 2.
- Fabry-Pérot / Grating Combination Observations (High spectral
resolution).
The wavelength is selected by setting the Fabry-Pérot unit to a
gap where one of the transmitting orders matches the correct wavelength.
The F-P unit is then scanned in steps of about
1/4 F-P resolution element.
The grating is used as an order sorter, with the maximum of the grating
transmission tuned to the desired wavelength so that unwanted F-P orders are
suppressed. Because the two detectors of a F-P detector  pair
look at slightly different grating resolution elements, this tuning can 
normally be done only for one detector. The other detector will not 
simultaneously deliver a useful signal as use of this redundant detector would 
require commanding of a slightly different grating position.
Figure 4.1 gives an overview of the wavelength ranges that could 
be simultaneously covered by the SW and LW or FP detector systems.
Figure 4.1:
Overview of possibilities for simultaneous use of LW grating
and SW grating (top) and FP and SW grating (bottom). Only combinations
within the hatched area were allowed, and simultaneous observations through
aperture 3 when observing with the FP were not allowed. The corresponding AOT 
band designations and aperture numbers are indicated.
|  | 
 
During an AOT observation, SWS scans the source spectrum by rotating the 
scanning mirror of the grating or varying the gap of the 
Fabry-Pérot. Each AOT may perform one or more exposures 
with an internal calibrator for a photometric check 
and a number of dark current measurements with the aperture 
closed. Hence, the data produced by any single AOT is generally of several 
different types.
Four AOT's, listed below, are defined for SWS. Operating modes have 
changed since the start of PV, and any such changes are noted in this document.
- -
- AOT 1: a single up-down scan 
for each aperture. It has four possible scan speeds and is primarily intended 
to quickly scan the entire SWS wavelength range at degraded 
resolution.
 
- -
- AOT 2: slow short up-down scans, designed to measure line profiles.
 
- -
- AOT 6: medium length grating scans with various options. Programmed as long up-down scan, possibly with 
reference scan. Used to observe long wavelength regions 
 
- -
- AOT 7: LW section devoted to FP observations. The 
SW section does almost the same as in AOT 6.  In this AOT priority is given to 
the FP.
Tables 4.1 and 4.2 summarise their properties. In 
all of these AOT's the measurements of photometric check, dark current and 
science data are interleaved in the ERD and SPD, while 
the AAR only contains science detector data for 
which an unique wavelength could be assigned to the light falling on that 
detector.
Table 4.1:
Various SWS AOT options
| AOT # | scan direction | reference scan | 
| 1 | up-down | no | 
| 2 | up-down | no | 
| 6 | up-down | possible | 
| 7  LW/FP | up | no | 
| SW/grating | up-down | possible | 
 
 
Table 4.2:
Possible combinations of reset interval tr, 
step size Ls in scanner steps, and number of scans nscan for each of 
the SWS AOT's
| AOT | Band |  |  |  |  |  |  |  |  | notes | 
| SWS01 | all | tr | 1 | 2 | 2 | 2 |  |  |  |  | 
|  |  | Ls | 32 | 32 | 16 | 8 |  |  |  | a | 
|  |  | nscan | 2 | 2 | 2 | 2 |  |  |  | 1 up, 1 down | 
|  |  | nint | 24 | 24 | 24 | 24 |  |  |  | b | 
|  |  | resolution | R/8 | R/8 | R/4 | R/2 |  |  |  | c | 
|  |  | duration | 1172 | 1944 | 3846 | 6570 |  |  |  | seconds | 
| SWS02 | 1 | tr | 1 | 2 | 4 | 4 |  |  | .. |  | 
|  |  | Ls | 1 | 1 | 1 | 1 |  |  | .. |  | 
|  |  | nscan | 2 | 2 | 2 | 4 |  |  | .. | up, down, etc. | 
|  |  | nint | 110 | 110 | 110 | 110 |  |  | .. | b | 
|  |  | duration | 1.3 | 2.7 | 5.4 | 10.8 |  |  | .. | minutes | 
| SWS02 | 2 | tr | 1 | 2 | 2 |  |  |  | .. |  | 
|  | 3 | Ls | 1 | 1 | 1 |  |  |  | .. |  | 
|  |  | nscan | 2 | 2 | 4 |  |  |  | .. | up, down, etc. | 
|  |  | nint | 110 | 110 | 110 |  |  |  | .. | b | 
|  |  | duration | 1.3 | 2.7 | 5.4 |  |  |  | .. | minutes | 
| SWS02 | 4 | tr | 1 | 2 | 2 |  |  |  | .. |  | 
|  |  | Ls | 2 | 2 | 2 |  |  |  | .. |  | 
|  |  | nscan | 2 | 2 | 4 |  |  |  | .. | up, down, etc. | 
|  |  | nint | 90 | 90 | 90 |  |  |  | .. |  | 
|  |  | duration | 1.3 | 2.7 | 5.4 |  |  |  | .. | minutes | 
| SWS07 F-P | 5-6 | tr | 1 | 2 |  |  |  |  | .. |  | 
|  |  | Ls |  |  |  |  |  |  |  | 1/4  | 
|  |  | nscan | 2 | 2 |  |  |  |  | .. | unidirectional | 
|  |  |  |  |  |  |  |  |  |  | 3 times or more | 
| SWS06 | 1 | tr | 1 | 2 | 2 | 4 | 4 | 4 | .. |  | 
| + |  | Ls | 4 | 4 | 2 | 2 | 1 | 1 | .. |  | 
| SWS07 |  | nscan | 2 | 2 | 2 | 2 | 2 | 4 | .. | up, down, etc | 
| SW section |  | nint | 27 | 27 | 55 | 55 | 110 | 220 | .. |  | 
| SWS06 | 2-3 | tr | 1 | 2 | 2 | 2 | 2 | .. |  |  | 
| + |  | Ls | 4 | 4 | 2 | 1 | 1 | .. |  |  | 
| SWS07 |  | nscan | 2 | 2 | 2 | 2 | 4 | .. |  | up, down, etc | 
| SW section |  | nint | 27 | 27 | 55 | 110 | 220 | .. |  |  | 
| SWS06 | 4 | tr | 1 | 2 | 2 | 2 | 2 | 2 | .. |  | 
|  |  | Ls | 6 | 6 | 3 | 2 | 1 | 1 | .. |  | 
|  |  | nscan | 2 | 2 | 2 | 2 | 2 | 4 | .. | up, down, etc | 
|  |  | nint | 30 | 30 | 60 | 90 | 180 | 360 | .. |  | 
 
Notes:
- a.
- Effectively. Actually 8 steps per sec.
- b.
- nint, the number of integrations per resolution element. This 
is the product of nscan, the number of detectors in an array (12), and 
the ratio between resolution and stepsize.
- c.
- Approximate effective spectral resolution cf AOT 2 - see section 
4.2.4.
Example output from all AOTs are given in sections 4.2 to 
4.5.
In all cases AOTs were actually commanded as sets of ICSs, 
Instantiated Command Sequences. These are more basic instrument commands to 
open the shutter, start the grating moving etc. The eight ICSs defined for SWS 
are listed in table 4.3. The use of these ICSs is described in the 
following sections.
Table 4.3:
ICSs
| ICS | Use | 
| SS0001 | Diffuse cal (not used) | 
| SS0002 | Grating cal (dark - on low - wait) | 
| SS0004 | SW grating scan for AOT 7 | 
| SS0005 | SW/LW grating scan for AOTs 1, 2 & 6 | 
| SS0006 | Dark and Wait for AOTs 1 & 2 | 
| SS0007 | Reset ICS | 
| SS0008 | Reset ICS | 
| SS0009 | FP LW | 
 
 
 
 
 
 
 
 
 
 
 
 Next: 4.2 AOT 1
Up: 4. AOT overview
 Previous: 4. AOT overview
SWS Instrument & Data Manual, Issue 1.0, SAI/98-095/Dc