The SWS Fabry-Pérot section may be used at wavelengths
between 11.4 and 44.5.
Instrument levels tests showed that performance
in the extended wavelength range outside the nominal design wavelength range of
15-35
was good enough to make the extended range available. A few
remaining peculiarities of the extended range are discussed together with the
following performance description. Results throughout this section refer to
use of the Fabry-Pérot AOT bands of Table 3.2
that have been selected to give the best tradeoff between the various design
goals (high sensitivity, low leakage etc.).
![]() |
The Fabry-Pérot spectral resolution depends on precise parallelity of the F-P plates. An in-orbit spectral resolution very close to Figure 5.1 was obtained after execution of the in-orbit parallelisation procedure of the SWS F-P.
The F-P sensitivity was characterized in instrument levels tests by doing
measurements on a calibrated blackbody source within the test cryostat. Fig.
5.2 shows the SWS Fabry-Pérot spectral
response. The conversion between units applicable to
astronomical sources (Flux density of a point source in Jy) and the resulting
instrumental signal (V/s) is given as a function of wavelength.
Diffraction losses caused by the fact that the
aperture is comparable in size to the ISO diffraction disk have been taken
into account; the flux used to derive Fig. 5.2 is the true
point source flux. The effect of these diffraction losses is however not large
(Fig. 5.3).
![]() |
The spectral response curve has structure on various wavelength scales.
Large scale trends are determined by the counteracting effects of detector
spectral response increasing with wavelength until close to the detector
cut-off
wavelength, and decreasing transmission of the Fabry-Pérot interferometer.
Structures on scales of microns or tenths of microns can be traced to the
detector spectral response curves and filter transmission curves. A rapid
low-amplitude modulation in the 11.4-16
range (not resolved in
Fig. 5.2) is caused by reflections between the surfaces
of the transmission filter used behind slit 1 in the long-wavelength
section of SWS. This `parasitic Fabry-Pérot' effect
is also observed for grating data using the same slit. The stronger
modulation detected at
in the long-wavelength
extended range is again caused by a parasitic Fabry-Pérot effect, this
time in a reflecting CaF2 filter. The nearby reststrahlen resonance
of CaF2 causes strong variations with wavelength of the indices of
refraction and absorption, leading to the observed modulation pattern. The
modulations were characterized with sufficient accuracy to allow good flux
calibration in the extended wavelength ranges.
![]() |
For sensitivity calculations, this may be summarized as
Band | 5 | 6 |
detector | FP-Si:Sb | FP-Ge:Be |
range (![]() |
11.4-26 | 26-44.5 |
![]() |
0.63 | 0.89 |
![]() |
1.00 | 3.00 |
![]() |
0.18 | 0.13 |
No saturation problems are expected for the SWS
Fabry-Pérot detectors. The saturation limit is given by the fact that the
integration ramp for 1s detector reset interval does not
hit the saturation level of the warm electronics for signals less than about
20,000 V/s. Although not foreseen in the current design of the offline
data processing, it is in principle possible to obtain data for signals about
seven times stronger by analysis of partial integration ramps. It is assumed
here that the expected strong flux has been correctly specified in the AOT
input, since only then the gain and reset time are set properly.
Taking typical values from Fig. 5.1 and
Fig. 5.2, the 20,000
V/s limit corresponds to a
saturating flux density of about 2 x 106 Jy, or a saturating flux in
an unresolved line of about 10-11 Wm-2.
Leakage has been quantified by analysis of the grating instrumental profile as seen by the F-P detectors in scans of solid state laser lines. Figure 5.5 shows the leakage as a function of wavelength. It must be considered preliminary, since it is based on extrapolating the grating instrumental profile from measurements only at a few wavelengths of laser lines.
The spectral response data in section 5.2.2 have not been corrected for the effects of leakage, i.e. they refer to the total signal.
![]() |
The amplitude of this `noise' depends on various factors like quality of grating wavelength calibration, ratio between grating stepsize and line profile FWHM, pointing stability. It is hence difficult to predict for an observation at a certain wavelength. Instrument level tests with optimized grating wavelength calibration showed `signal-to-noise' around 200. In the case of an observation of a single line on a smooth continuum, part of the `noise' could be calibrated out by observing a larger region of continuum around the line, provided the non-systematic contribution of pointing jitter is not too large.
![]() |
![]() |