next up previous contents index
Next: 7.6 Wavelength calibration accuracy Up: 7. SWS Calibration Previous: 7.4 Flux calibration accuracy


7.5 Wavelength calibration

The astronomical quantity of interest is an array of flux as a function of wavelength. Hence, a relationship must be established between each grating scanner position (measured by a Linear Voltage Differential Transducer (LVDT) and colloquially given the units `LVDT') as contained in the ERD and its corresponding wavelength. The relation is calculated in two steps: first the scanner position is converted to a scanner angle using a calibration file. Then the following equation can be applied to calculate the wavelength of light falling on the detectors:


\begin{displaymath}
\frac{\lambda\times N}{D} =
\sin(\theta + \beta) + \sin (\theta + \delta)
\end{displaymath} (7.4)

where,
$\beta$
is the angle from the optical axis to the entrance or light source used, as seen from the collimator,
$\delta$
is the angle from the optical axis to the detector pixel,
$\theta$
is the angle of incidence on the grating for a beam entering along the optical axis for the current scanner position,
D
is the grating constant,
N
is the spectral order.

$\beta, \delta$ and D are constants, taken from calibration file 16. The spectral order N is determined by calculating all the orders that give wavelengths consistent with the transmission limits of the band/order/aperture/filter combination. For valid observations only one order should give a wavelength that satisfies this criterion.

The wavelength range - band/order/aperture/filter combination relationship is shown in Table 3.1.

For the Fabry-Pérot, first the grating wavelength is calculated. Calibration file 12 then transforms the nominal FP-position (an engineering parameter of no interest to the observer) into a FP-gap. This mechanical gap width must be corrected to get the optical gap width. The wavelength which passes through the FP is the one of which a whole multiple fits exactly in two times the optical gap width.


next up previous contents index
Next: 7.6 Wavelength calibration accuracy Up: 7. SWS Calibration Previous: 7.4 Flux calibration accuracy
SWS Instrument & Data Manual, Issue 1.0, SAI/98-095/Dc